30
Measured Data Header Table
Characters
Sent
Range of
x...x
Meaning
Notes
Fx…x
100,000 to
200,000,000
Center frequency in
Hz
These numbers depend on
the version
Wx…x
0 to
51.2MHz or
64MHz
Sweep width in Hz
Max width depends on plot
width of 80 or 100 points
Nx…x
1 to 200
Number of data points
Data sent in pairs (Z,A or
R,X etc) count as one data
point
Dxxx
101 to 104
Data format
See D in unit setup
command. Defines the type
of data to follow
Measured Data Formats
Characters
Sent
Range of
x...x
Range of
y...y
Meaning
Notes
Rx..xXy..y
0 to
32,767
-32,768 to
32,767
R = 10x ohms
X = 10x j ohms
insert ‘.’ Before last digit.
Resist. & React. D = 101.
Zx..xAy..y
0 to
32,767
-899 to
899
Z = 10x ohms
A =10x degrees
insert ‘.’ Before last digit.
Total Z & Z angle D=102.
Vx..xLy..y
100 to
10,000
0 to
10,000
V = 100x
(V)SWR
L=100xRtn loss
dB
insert ‘.’ Before last 2
digits. VSR & Rtn Loss
D=103
Mx..xDy..y 0 to 100
0 to 3600
M = 100x RC
mag.
D = 10x RC
angle
Reflection coefficient or
Smith Chart vector, insert
‘.’ in appropriate place
D=104
9.3.1.2.2 Q
Returns the letter Q with a number equal to 10 times the calculated Q value. The
last character sent is the EOS marker
9.3.1.2.3 D
Returns the letter D with a number equal to the measured length in millimeters.
The last character is the EOS marker.