29
9.3.1.1.3 D
Send the distance reading in millimeters. This command only operates when the Bravo
MRI II unit is in the cable analyzer mode. . No EOS marker is required, but one may be sent
if desired.
9.3.1.1.4 Mxx
Read the data stored in memory location xx. The EEPROM memory location contains
the unit setup info, and sometimes the plot from a single sweep saved by the user. The
maximum value of xx depends on the unit version, and can be obtained using the S000
command. The number sent for xx should always be a 2 digit number, for values less than
10 a leading zero is required.
Location 00, last configuration used/power up default.
Locations 01 to 16, user defined instrument presets, no data.
Locations 17 to 24: presets and plot data.
This command requires no EOS marker, but one may be sent if desired.
9.3.1.1.5 S000
Read the unit’s current setup. The entire setup is read in a fixed format block of data.
This command requires no EOS marker, but one may be sent if desired.
9.3.1.2 Data Request Responses
9.3.1.2.1 Measured Data Response (R)
The unit response to an ‘R’ command will send a common header followed by a block of
data, the contents of the data are determined by the instrument state established by the
keypad or previous setup commands. The information sent in the header is shown in the first
column of the table. The header and data are sent in one continuous string, but they are
divided up in the tables for clarity. The numbers sent will always be integers (0 to very large)
and do not contain decimal points or commas.