(1) When the counter is set to Remote, the RMT lamp to the left of the
front panel readout goes on.
(2) While the RMT lamp is on, the following switches are disabled:
FUNCTION selector switch
GATE TIME selector switch
RESET switch
HOLD switch
MASK switch
BURST switch
All switches on the TR1644 Calculation Unit (accessory)
* The sliding switches and controls at the input section remain valid in
the Remote State.
(3) The Remote settings (such as function) are retained when the state of
the counter is changed from Remote to Local by the controller unless
the POWER switch o f the counter is set to OFF. When the counter is
returned to the Remote state after altering the state to Local and chang
ing the settings by operating the panel switches, the previous Remote
settings are retained. (It appears that the settings such as function are
also changed by simply changing the state from Local to Remote.)
(See the positions indicated by the arrows in Figure 4-4.)
4-5-2. Panel Description
4 - 1 1
Summary of Contents for TR5821
Page 19: ...A d a p t e r Fig 3 1 Power cable plug and adapter ...
Page 23: ...7Inm 9 8 8 3 8 6 9 Fig 3 2 Panel description 3 6 ...
Page 26: ...3 9 ...
Page 54: ...4 6 Program m ing N otes 4 6 1 Sim plified O peratio nal F low C hart D ata Sending 4 15 ...
Page 56: ... 2 When s e r i a lpollingi snot used Addressed to talk 4 17 ...
Page 82: ...7 3 ...
Page 100: ...D im ensions of chassis 2 ...
Page 103: ...SIDE VIEW Ut TR5821 EXTERNAL VIEW ...
Page 104: ...F R O N T VIEW REAR VIEW ...
Page 105: ...4 280 u TR5822 EXTERNAL VIEV ...
Page 106: ...FRONT VIEW R EAR VIEW ...
Page 107: ...SIDE VIEW U nit TR5823 EXTERNAL VIEW ...
Page 108: ...101 75 tOaOOQQ Oa QC QOj 5 kS a a a a o o o o a a a y TOP VIEW FRONT VIEW 102 ...