(7 ) MASK
The signal to be measured is masked by pressing
1 to light the lamp
within the switch. Set the masking time with control d ). (FREQ. A can
not be masked.)
® MASK TIME
A control for setting the masking time.
@ SLOPE
Trigger point slope selector switch. + sets the trigger point at the positive
slope and — sets it at the negative slope.
@ LEVEL
A control for triggering the signal to be measured at the proper level.
The trigger voltage can be adjusted within a range of about —1 V to
+1 V.
(H) SENS.
Selects the input sensitivity.
@ FILTER
When the switch is set ON, the 10 0 kHz (approx.) low-pass filter is
activated on both A and B channels.
(13) LEVEL
See @ .
@ AC-AUTO-DC
Input coupling selector switch. AC and AUTO cut out the DC compo
nent and routes the AC component into the input circuit. AUTO auto
matically sets the trigger voltage to the 50% level of the voltage to be
measured. DC routes both AC and DC components into the input
circuit.
(0 ) INPUT B
Channel B input connector.
@
SEP./COM. A
SEP. is used for two separate inputs A and B, whereas with COM. A the
signal to be measured on channel A is common to both channels.
(g ) SLOPE
© SENS.
See (H).
Summary of Contents for TR5821
Page 19: ...A d a p t e r Fig 3 1 Power cable plug and adapter ...
Page 23: ...7Inm 9 8 8 3 8 6 9 Fig 3 2 Panel description 3 6 ...
Page 26: ...3 9 ...
Page 54: ...4 6 Program m ing N otes 4 6 1 Sim plified O peratio nal F low C hart D ata Sending 4 15 ...
Page 56: ... 2 When s e r i a lpollingi snot used Addressed to talk 4 17 ...
Page 82: ...7 3 ...
Page 100: ...D im ensions of chassis 2 ...
Page 103: ...SIDE VIEW Ut TR5821 EXTERNAL VIEW ...
Page 104: ...F R O N T VIEW REAR VIEW ...
Page 105: ...4 280 u TR5822 EXTERNAL VIEV ...
Page 106: ...FRONT VIEW R EAR VIEW ...
Page 107: ...SIDE VIEW U nit TR5823 EXTERNAL VIEW ...
Page 108: ...101 75 tOaOOQQ Oa QC QOj 5 kS a a a a o o o o a a a y TOP VIEW FRONT VIEW 102 ...