Timer
The maximum test run-time is 59 minutes and 59 seconds. To
program the test run time, touch the green button below the “TEST
TIME” box.
Program the minutes and seconds for your test, then touch the enter
key.
The test time length will display in green under the “TEST TIME”
button. The time remaining is displayed below that in yellow.
Light
The interior light may be turned on or off by tapping the “LIGHT”
button.
Vertical and Horizontal Tappers
The tapping mechanisms on the VariSifter™ assist in reorienting
particles on the sieve surface and breaking up agglomerates. Each
may be independently turned on or off, per user preference. The
respective buttons will be green if the function is on and grey if the
function has been turned off.
NOTE: When sifting powders 45µm and finer, addition of the
horizontal tappers may be beneficial. Use of the vertical and/or
horizontal tappers may not necessarily help with static electric charge
issues.
Removing the Stack Assembly
The stack assembly can be removed from the test chamber by first
removing the Column Stop in front of the test stack. Grasp the locking
arms on the column lock assembly and pull straight down while
squeezing the arms together. This will lock the assembly together so
the stack can be slid straight out of the test chamber. WARNING: Do
not leave the column lock arms clamped while the test stack is out of