Appendix E
OsmoTECH XT LIS Specification
86
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
Record field contents
In the examples below, the overhead characters of
<STX>
and frame number at the front and checksum
plus
<CR><LF>
at the end are not shown.
Optional fields and UTF8-encoded fields are indicated.
E.6 Header record
The fields used in this record are as specified in LIS2-A2:
6.1
Record type ID—
H
identifies the record as a
message header record.
6.2
Delimiter definition—The five Latin-1 characters
immediately following the
H
record type ID define
the delimiters used throughout the subsequent
records of the message.
–
The second character in the header record is
the
field delimiter
.
–
The third character is the
repeat delimiter
.
–
The fourth character is the
component
delimiter
.
–
The fifth character is the
escape character
.
6.5
Sender ID—This text field defines the sender. In
OsmoTECH XT osmometers, this is defined by
the instrument type and the software version.
6.10
Receiver—Text field containing the name of the
receiver. This field is used to verify that the
transmission is for the receiver. In the instrument,
this text is the word
LIS
.
6.12
Processing ID—Message processing where
P
= production (sample test data) and
Q
= quality
control.
Example
H|\^&|||OsmoTECHXT^V1.0|||||LIS||P|LIS2-
A2|20161110082005<CR><ETX>
E.7 Sample information record
The fields used in this record are as specified in LIS2-A2:
7.1
Record type ID—
P
identifies the record as a
sample information record.
7.2
Sequence number—Sequence number (integer)
of the sample record within the message.
7.3
Practice-assigned ID (optional)—Text field
containing the unique sample ID assigned by the
practice.
7.4
Laboratory-assigned ID (optional)—Text field
containing a unique sample ID assigned by the
laboratory.
Example
P|1|PracticeID|LabID<CR><ETX>
E.8 Test order record
The fields used in this record are as specified in LIS2-A2:
8.4.1
Record type ID—
O
identifies the record as a test
order record.
8.4.2
Sequence number—Number (integer) of the
order record for the sample.
8.4.3
Specimen ID (optional, UTF8-encoded)—Text
field containing the unique specimen ID assigned
by the LIS.
If no
instrument specimen ID
(8.4.4) is specified,
the specimen ID
must
be specified.
8.4.4
Instrument specimen ID (optional)—Text field
containing the unique specimen ID assigned by
the instrument either from reading a barcode or
as entered by an operator.
If no
specimen ID
(8.4.3) is specified, the
instrument specimen ID
must
be specified.
8.4.5
Universal test ID—Text field containing up to four
parts describing the test. This code is always the
four-letter code
OSMO
for the instrument.
8.4.6
Priority flag (optional)—Single character field
indicating the priority of the test. The only codes
recognized by the instrument are
S
for STAT and
R
for routine (default).
8.4.7
Requested/ordered date and time (optional)—
Text field containing the date/time of the order. If
the date/time is in the future, the system does
not look for the specimen until after that date/
time.
8.4.8
Specimen collection date and time (optional)—
Text field indicating the date/time that the
specimen was collected.
8.4.10
Collection volume (optional)—Text field indicating
the volume of the specimen. If no unit is specified
(with a component delimiter), the unit defaults to
milliliters.
8.4.11
Collector ID (optional)—Text field indicating the
ID of the specimen collector.