REJECT AND SIGNAL DEVICES TIMING AND OPTIONS
A brief description of each reject term is shown below.
The “REJECT OPTIONS” menu can be reached by MENU / LEVEL / PASSWORD / REJECT OPTIONS.
Level 2 or 3 gives access to reject settings.
The reject menu appears below. Please note there are 2 pages:
Reject delay time
- Sets the time between detection of metal and the activation of the reject device. It can be
adjusted in tenths of a second from 0-30 seconds.
Reject operate time
- Sets the time that the reject device stays activated. It can be adjusted in tenths of a second
from 0-30 seconds.
On detection of metal, the reject device, light, and or alarm will activate after the "delay" time has elapsed. The
reject and signal devices will continue to operate until the "operate" time has elapsed.
Similarly, a conveyor that has been wired to stop upon detection will do so after the "delay" time has elapsed.
Unlike rejects, however, the conveyor must be restarted by depressing the start button on the conveyor's motor
starter. The conveyor will not start until the metal detector operate time has elapsed.