32
Operation
Figure 3-6: Thermocouple Connection
3.3.5
RTD Input Mode
The resistance temperature detector (RTD) measures tempera-
ture by correlating the resistance of the RTD element with temper-
ature. The USB-2401 can generate a steady 0.5 mA excitation
current source to each channel in RTD input mode to measure
cross-voltage on the RTD. The actual input range is ±2.5V with a
formula of voltage to RTD resistance conversion of:
Since the excitation current can only drive cross-voltage up to
1.5V with good linearity, the maximum equivalent value of the RTD
resistor is limited to 3k
.
The USB-2401 can support two, three, and four-wire RTD mea-
surement. Adopting three- and four-wire connections rather than
two-wire can eliminate connection lead resistance effects from
measurement. Three-wire connection is sufficient for most pur-
poses and most universal industrial applications. Four-wire con-
nections are used for the most precise application requirements.
ADC
MUX
SC+
SC-
Thermocouple
USB-2401
0.0005
V(volt)
)
(
=
Ω
RTD
Summary of Contents for USB-2401
Page 8: ...viii List of Figures This page intentionally left blank...
Page 10: ...x List of Tables This page intentionally left blank...
Page 22: ...12 Introduction Figure 1 3 USB 2401 Module Side View...
Page 23: ...Introduction 13 USB 2401 Figure 1 4 USB 2401 Module Front View 41 3 114...
Page 25: ...Introduction 15 USB 2401 Figure 1 7 Module In Stand Front View...
Page 26: ...16 Introduction Figure 1 8 Module Stand Top View 20 4 20 4 B 26...
Page 30: ...20 Introduction This page intentionally left blank...
Page 34: ...24 Getting Started Figure 2 4 Module Pre Rail Mounting Figure 2 5 Module Rail Mounted...
Page 36: ...26 Getting Started This page intentionally left blank...
Page 56: ...46 Calibration This page intentionally left blank...