Cleaning and Maintenance 

6.1

 

Cleaning 

Before cleaning make sure the scale has been disconnected from the socket. For cleaning just 
use a damp cloth or common disinfection products. Do not use any aggressive cleaning fluids. 
Models without rechargeable battery can be cleaned under running water.  
 
 

Please make sure before starting cleaning, if there’s a rechargeable battery inside.

 

Otherwise the scale or the rechargeable battery can be damaged. 

If a rechargeable battery is inside, make sure no water or cleaning fluid penetrates 
the scale. 

It isn’t allowed to clean the scale with a high-pressure cleaner! 

6.2

 

Storage and Transport 

Keep all original parts and pieces for an eventual sending back of the scale in order to avoid any 
possible damage during transport which automatically leads to no warranty. 

6.3

 

Disposal of Waste 

Never dispose the scale at the end of life to domestic waste. 

Batteries  and  rechargeable  batteries  are  not  allowed  in  domestic  waste  and  have 
to be disposed fully discharged or with insulated poles. Otherwise there’s a danger 
of an electrical short! 

Please ask at your local administration for an environmentally correct disposal. 

6.4

 

Warranty 

A  one  year  warranty  period  from  date  of  purchase  will  be  guaranteed  for  either  repair  or 
replacement  once  it  is  proven  that  the  defects  are  caused  by  poor  material  or  poor 
workmanship  (please  keep  the  receipt  for  presentation).  All  moveable  parts  such  as  batteries, 
cables, plug-in power supply units, rechargeable batteries etc. are excluded. Warranty does not 
apply to normal appearance of usage or sign of wear or any damages caused by accidents or 
misuse.  Products  which  have  been  opened  by  unauthorized  persons  are  automatically  out  of 
warranty. 

Customers  abroad  are  requested  to  contact  their  local  dealers  in  the  event  of  any  warranty 
becoming applicable. 

6.5

 

Safety Instructions 

Do NOT use the scale

 

If the mains plug or cable shows visible damages 

In this case contact your local dealer or service partner. 
 

Summary of Contents for KWE Series

Page 1: ...ss Steel Scales Compact Scale in Stainless Steel Housing Series KWE KWE 190107 Rev008 UM en ADE Germany GmbH Co KG Hammer Steindamm 27 29 22089 Hamburg GERMANY 49 0 40 432776 0 Fax 49 0 40 432776 10 E mail info ade hamburg de ...

Page 2: ...11 Check weighing Minimum and maximum values with acoustical alarm 5 2 12 Counting Function 6 3 Environment Parameter Setting 6 3 1 Adjustment display precision 7 3 2 Weighing unit kg g lbs 7 3 3 Zero tracking 7 3 4 Stability Filter 7 3 5 Power Saving Mode Standby only for models with rechargeable batteries 8 3 6 Display Update Speed 8 3 7 Creep Tracking Range 8 3 8 Printer setting 9 4 Display Sym...

Page 3: ...o that there can be stored minimum and maximum values combined with an acoustical signal in the scale e g for weighing always the same weights The stainless steel housing allows using the scale in humid working environments The scale is a precision instrument and must be treated accordingly Therefore kindly read and follow the Operating Manual in all respects 2 1 Check carton contents Compact Scal...

Page 4: ...d on the adapter 2 4 Power immediate operation Working with mains all three LCD High Mid Low are on For models with rechargeable batteries All three segments lighting Power voltage of rechargeable batteries is more than 6 3V Two segments lighting Power voltage of rechargeable batteries is between 6V und 6 3V One segment lighting Power voltage of rechargeable batteries is under 6V 2 5 Power Charge ...

Page 5: ...age of the rechargeable batteries is less than 5 6V the scales turn off automatically 2 7 Zero Operation When scale is on ON is the key to zero It is only available within 4 of the maximum capacity otherwise it will immediately display HHHHHH 2 8 Displaying the voltage When scale is on it will display the voltage when pressing ON key for 3 seconds Now if press ON again the scale will return to wei...

Page 6: ...ET to change the digit press TARE to change the digit value and press ON to confirm Then the setting of minimum value is done and the arrow at Over is lighting Use the same way to finish the setting of the maximum value When finished setting it returns to weighing mode and the weighing alarm function is on If the weight is upper the maximum value the Over indicator is lighting while arrow to Under...

Page 7: ...n memory when scale is switched off For returning to weighing mode press both SET and ON together for 3 seconds and choose pcs_n by pressing TARE Confirm your choice with ON The setting will be stored and scale return to weighing mode When scale is switched on next time it starts in weighing mode Piece counting function doesn t work if unit lbs is selected 3 Environment Parameter Setting The user ...

Page 8: ...isplay In this menu the zero tracking can be adjusted if there are weighing performances with weight increasing which are very slow There can be chosen between 11 adjustments 0 5 in half steps by pressing TARE again and again Default is Zer_2 Remark If you have selected a higher graduation PrE 3 1 Adjustment display precision select a high value for stability filter 3 4 Stability Filter Please sta...

Page 9: ...lay is showing 0 backlight automatically switches off By putting something on the weighing platform again backlight of display will switch on again If nothing will be put on the weighing platform within ten minutes the scale automatically switches off 3 6 Display Update Speed Please start as described in the beginning of this chapter to enter this menu and SPd is shown in the display By pressing T...

Page 10: ...hargeable battery has been charged fully _Inlt_ Faulty weighing 5 Display Symbols Fault Cause Solution Faulty weighing result Parameter of load cells have changed Re calibration Display shows HHHHHHHH Over zero range Re calibration Load cell or PCB damaged Display shows LLLLLLLL Load cell or PCB damaged Rear display doesn t show anything Parameter fault Correction of setting as shown in the manual...

Page 11: ...omestic waste Batteries and rechargeable batteries are not allowed in domestic waste and have to be disposed fully discharged or with insulated poles Otherwise there s a danger of an electrical short Please ask at your local administration for an environmentally correct disposal 6 4 Warranty A one year warranty period from date of purchase will be guaranteed for either repair or replacement once i...

Page 12: ... EU regarding electromagnetic compatibility RoHS Directive 2011 65 EG as appropriate amended This declaration loses their validity in case modifications on the scales are made without manufactures approval Hamburg January 2019 ADE Germany GmbH Co KG Hammer Steindamm 27 29 22089 Hamburg Germany 7 Declaration of Conformity Manufacturer ADE Germany GmbH Co KG Hammer Steindamm 27 29 D 22089 Hamburg Ge...

Page 13: ...115 mm Dead Weight 4 1 kg 3 8 kg Connecting cable 1 00 m 2 00 m Operating Temperature 10 C to 40 C 10 C to 40 C Internal Protection IP65 IP68 Power Supply 230V mains 230V mains and rechargeable battery 6 3V 6V4Ah refitting with rechargeable battery not possible Display Large LCD Display with second display on backside height of digit 20mm Casing Stable stainless steel housing Weighing platform sta...

Page 14: ...13 KWE 190107 Rev008 UM en ...

Page 15: ...KWE 190107 Rev008 UM en 14 ...

Page 16: ...15 KWE 190107 Rev008 UM en Händler ...

Reviews: