background image

 

 

13. Do not remove the silicone rubber protective sleeve from the enclosure of equipment in Hazardous area 
14. IS parameters must be observed for different external measuring ports on the different models in accordance with 
Ex instruction strictly and completely 
15. Intrinsically safe electrical system between IS apparatus and Associated apparatus or other IS apparatus: 

Item 

I.S Interface 

External Measuring Instrument 

System 

Equipment group 

IIC 

IIC 

IIC 

Level of protection 

ia 

ia 

ia 

Temperature class 

T4 

T4 

T4 

Ambient temperature 

-20°C~+50°C 

-20°C~+50°C 

 

Voltage 

Uo 

Ui(30V) 

 

Current 

Io 

Ii(100mA) 

 

Power 

Po 

Pi(0.75W) 

 

Cable parameters 

 

 

 

Capacitance 

Co 

Ci 

Cc 

Inductance 

Lo 

Li 

Lc 

L

/

ratio 

Grounding  

Isolated 

Isolated 

Isolated 

Uo  

  Ui

 

Io    

  Ii

 

Summary of Contents for 260Ex

Page 1: ...1 260Ex Handheld Multichannel Reference Recorder...

Page 2: ...ADT260Ex Handheld Multichannel Reference Recorder User Manual Version 2211V01 Additel Corporation...

Page 3: ...s for the ADT260Ex Handheld Multichannel Reference Recorder To ensure correct operation and safety please follow the instructions in this manual Additel Corporation reserves the right to change the co...

Page 4: ...tion environmental temp 20 10 1 year accuracy 10 1 2 3 Simulate RTD measure accuracy environmental temp 20 10 11 1 3 Basic Structure 14 1 4 Standard package include 16 1 5 Power supply description 17...

Page 5: ...3 Switch measurement 27 3 3 4 Pulse measurement 27 3 3 5 Frequency measurement 27 3 4 Filtering 28 3 5 Scaling 29 4 System settings 31 4 1 Bluetooth communication 31 4 2 Power management 31 4 3 Syste...

Page 6: ...6 1 2 Multi channel list mode 38 6 2 Channel data 39 6 3 Channel curve 39 6 3 1 Following Mode 39 6 3 2 Eye pattern mode 39 6 4 Save the data 39 7 Applications 40 7 1 Differential pressure module 40 7...

Page 7: ...18 Figure 4 Control center 20 Figure 5 Calibrator main screen 23 Figure 6 RTD measurement 25 Figure 7 Voltage frequency pulse switch wiring 26 Figure 8 Current measurement 27 Figure 9 Filter settings...

Page 8: ...ct please check the appearance of the product for any damage If the product is damaged or malfunctions do not use it and contact Additel Do not touch the metal part of the probes or test cables during...

Page 9: ...ts of IEC 60664 1 and IEC 60664 3 by manufacturer 8 Tamb 20 to 50 9 USB port should only be used in Non hazardous areas 10 PX connections to be made with pressure measuring module CDPX Ex only the con...

Page 10: ...15 Intrinsically safe electrical system between IS apparatus and Associated apparatus or other IS apparatus Item I S Interface External Measuring Instrument System Equipment group IIC IIC IIC Level o...

Page 11: ...information for equipment name National regulations ATEX directive 2014 34 EU and IECEx scheme 02 Certificate No T V 20 ATEX 8509 X and IECEx TUR 20 0009X Ex Marking for EU Type Examination II 1 G Ex...

Page 12: ...i nF Li mH J1 J3 HART external power and external resistance 30 100 750 10 0 J2 J3 Current measurement 30 100 750 1 0 J4 J3 Voltage frequency and switch measurement 30 100 750 1 0 J1 J2 Loop Circuit T...

Page 13: ...witch measurement 5 0 85 1 1 87 100 J1 J2 Loop Circuit Transmitter Current Measuring HART Internal Power and Resistance 25 2 79 16 499 0 107 9 J5 J8 J9 J12 Resistance measuring 5 21 26 3 85 30 Intende...

Page 14: ...s display the trend curve and support Bluetooth communication Powerful logging capability and high level intrinsically safe make it widely used in ex proof pipeline field container pressure test air l...

Page 15: ...ospheric pressure sensor Barometric Accuracy 55Pa Measurement Rate mV V mA frequency and RTD 3 times sec Pressure module 1 10 times sec selectable 3 as default Barometric 1 times sec Data Storage Logg...

Page 16: ...p A B C and D T4 Class I Zone 0 AEx ia IIC T4 Ga Ta 20 Cto 50 C Ex ia IIC T4 Ga UKCA EX CE Certification TUV IEC61326 IEC61010 Rohs Compliance Rohs II Directive 2011 65 EU EN50581 2012 Protection Leve...

Page 17: ...Frequency Measurement 0 01 50000Hz auto range 0 005 RDG 2 last digit 6 digit auto resolution Min threshold voltage 2 5V Units Hz kHz MHz CPM CPH s ms us Switch On Off Measurement Inspection voltage 3...

Page 18: ...2 The accuracy of the 2 wire method is 50m more than the accuracy of the 4 wire method Note 4 Minimum threshold voltage for frequency and pulse measure 2 5V Note 5 The frequency measure supports foll...

Page 19: ...T100 3916 PT100 3926 200 850 200 200 0 10 200 600 0 16 600 850 0 20 PT200 385 200 850 200 200 0 32 200 300 0 34 300 600 0 41 600 850 0 48 PT400 385 200 850 200 0 0 15 0 200 0 18 200 600 0 25 600 850 0...

Page 20: ...13 Cu100 428 50 150 50 150 0 08 Ni100 617 Ni100 618 60 180 60 0 0 06 0 180 0 05 Ni120 672 80 260 80 260 0 05 Ni1000 50 150 50 150 0 07 Note It complies with the international temperature scale ITS90...

Page 21: ...14 1 3 Basic Structure Figure 1 Basic Structure...

Page 22: ...rating the device 3 Lemo port A Connect to external pressure module 4 Lemo port B Connect to external pressure module 5 Adaptor port Power supply from adaptor 6 USB slave port For USB communication 7...

Page 23: ...dited calibration certificate 1 pc Optional ADT158Ex Built in digital pressure module ADT161Ex External digital pressure module 9530 Additel ACal calibration software 9060 Pressure module connection c...

Page 24: ...Li ion battery Ex Proof Li ion battery is chargeable by external power independently The power adaptor is adaptable for standard in various countries Do not expose the battery to fire or short circui...

Page 25: ...for display and basic functions The ADT260Ex will automatically go to Data Logger function after powered on It can also be returned to main interface for all the functions Refer to Figure 3 Figure 3 M...

Page 26: ...cates the external module A is connected while indicates not connected 5 Module B icon indicates the external module B is connected while indicates not connected 6 Message center icon flashes means th...

Page 27: ...20 2 1 3 Main function guide The main function guide at the bottom of the interface provides access to three main functions of the device data logger data manager and setup Figure 4 Control center...

Page 28: ...time diagnostic hardware status of the device If some of the function fails check the cause here as in Figure 5 Click icon in the diagnostic center to see the historic abnormal log Click icon to take...

Page 29: ...el there is the current measured signal of this channel The icon of the signal type and the corresponding range information is displayed on the upper left Click the icon in the upper right corner to s...

Page 30: ...by internal pressure module external pressure module A B differential pressure and barometer The external pressure A and B are only available when the external pressure module is connected The differ...

Page 31: ...ription Resolution 4 5 6 Resolution of current measurement channel Pressure type GP AP Current measurement channel pressure type Measurement frequency 1 10Hz Data acquisition frequency of the current...

Page 32: ...measurement Click the Settings of the RTD measurement channel to set various parameters for the functions Table 7 RTD settings Subject Valid Value Description Sensor type RTD Select the type of therma...

Page 33: ...d adapt to more usage scenarios two different ranges can be selected for voltage measurement in this device 30 30 V and 300 300 mV the user should select the appropriate measurement item according to...

Page 34: ...he pulse measurement function is the same as the voltage measurement After wiring please switch the measurement signal channel to 0 9999999 Pulse measurement allows the user to set the pulse counting...

Page 35: ...filtering methods first order linear filtering and moving average filtering to process data to meet the needs of different scenarios Click the menu button of the measurement channel select the filter...

Page 36: ...disable the filter function Method first order linear filter or moving average filter Filtering method Filter coefficient 0 1 Only available when the first order linear filter is selected Number of s...

Page 37: ...30 Figure 10 Scaling configurations...

Page 38: ...ction and the query function of the Bluetooth device name and physical address 4 2 Power management Click the Power Management menu item in the system setting interface to enter the power management i...

Page 39: ...alibration interface to restore to the factory calibration data of the device Click the Cancel zero button to clear the zeroed data of the current signal 4 4 Services 4 4 1 Maintenance This function i...

Page 40: ...ation as in Table 10 Table 10 Sound settings Subject Valid Value Comment Sound enable On Off Sound enable setting Prompt sound On Off Prompt sound setting Touch sound On Off Touch sound setting Over r...

Page 41: ...hours enable 24 hour or 12 hour format 4 6 Product information Product information is read only information includes basic information and the module information Basic information including model seri...

Page 42: ...ged by functional modules and the data saved by each function is managed in the corresponding folder which is convenient for users to browse The functional modules that can save data files are snapsho...

Page 43: ...progress bar at the bottom and the percentage will show the progress and the time beside the progress bar will show the remaining time Click the stop icon on the upper right to terminate the current l...

Page 44: ...anual switch at present Click the left and right icon or the physical direction key to switch the channels When the top right area shows or it means automatic switch mode is activated and the device w...

Page 45: ...n figure 12 The channel can fit with the sizes and when logging more than 3 channels there will be the icon for page switching click the icon or the physical direction key to switch between pages Figu...

Page 46: ...rded on the curve For 1x following every recorded point will be added for one point on the curve For 2x following every two recorded points will be added for one point on the curve and for 4x followin...

Page 47: ...e Enable or disable the differential pressure module When the differential pressure module is enabled the differential pressure module channels will be displayed in the related functions channel list...

Page 48: ...tiple pressure units or temperature units for conversion Click the unit converter icon on the main interface of the device to enter the unit converter Click icon on the right top to select and switch...

Page 49: ...it after calculation 7 4 1 Leak Test Performing 1 Connect to air circuit a Connect the device to the air circuit b Click the start icon at the bottom of the screen apply pressure to the leak test poin...

Page 50: ...0 00 00 10 00 Set test time Pressure unit Different modules support different pressure units Select corresponding pressure unit 7 5 1 PSV test performing 1 Connect air circuit a Connect the device to...

Page 51: ...ADT260Ex provides a sensor library function In the sensor library function users can define new sensor types according to their needs and set relevant parameters in the sensor The sensor library supp...

Page 52: ...45...

Reviews: