42
the pressure module column to set the test parameters, see the table below:
Table 13 Leak test setting
Subject
Valid value
Description
Pressure type
GP/ AP
The pressure type being recorded, view detail according to
device’s module
Pressure unit
Pressure unit
Select corresponding pressure unit
Waiting time
Numbers
The pressurizing time of external module
Test time
Numbers
Total record time
Unit of the rate
Second/ minute
Displayed unit after calculation
7.4.1 Leak Test Performing
1. Connect to air circuit
a. Connect the device to the air circuit;
b. Click the start icon at the bottom of the screen, apply pressure to the leak test point;
c. Start counting down the waiting time, and record the real-time pressure at the end of the waiting time as the initial
pressure;
d. Start the test, count down the test time, and start to calculate the real-time leakage, real-time leakage = initial
pressure-real time pressure;
e. Record the pressure at the end of the test time as the end pressure;
f. The entire leak detection process ends, and the final leakage = initial pressure - end pressure;