6240B DC Voltage Current Source/Monitor Operation Manual
9.1 Source and Measurement
9-7
Settling time:
Time to reach the final value
0.1% when varying from zero to the full scale.
Setting conditions: Source values and limit values are full-scale settings.
Load conditions:
Pure resistance load, and load capacitance of 200 pF or less
Typical value
Source
range
Limiter
range
Settling time
Output response
FAST
SLOW
Voltage source
(Output current: 4 A
1 A in the 15 V range)
300 mV
4 A
200
s or less
1 ms or less
3 V
100
s
or less
300
s or less
15 V
300
s
or less
700
s or less
Current source
(Output voltage: 15 V
10 V in the 4 A range)
30
A
15 V
1.5 ms or less
2 ms or less
300
A
400
s or less
700
s or less
3 mA
30 mA
300 mA
1 A
1 ms or less
2 ms or less
4 A
450
s or less
700
s or less
Typical value
Source
range
Limiter
range
Settling time
Output response
FAST
SLOW
Voltage source
(Output current: 20 % or less of full sale)
300 mV
3 mA to
300 mA
50
s or less
200
s or less
3 V
30
s or less
100
s or less
15 V
100
s or less
300
s or less
Current source
(Output voltage: 1 V)
300
A
3 V
50
s or less
100
s or less
3 mA
30 mA
300 mA
1 A
100
s or less
200
s or less
4 A
50
s or less
150