6240B DC Voltage Current Source/Monitor Operation Manual
9.4 General Specifications
9-12
9.4
General Specifications
Operating environment conditions: Temperature 0 ºC to +50 ºC, relative humidity 85 % or below with no
condensation
Storage environment conditions:
Temperature -25 ºC to +70 ºC, relative humidity 85 % or below with
no condensation
Warming up time:
60 minutes or longer (until it settles in the specified accuracies.)
Display:
16 segments
12 digits vacuum fluorescent display
Power supply:
AC power 100 V, 120 V, 220 V, and 240 V (User selectable)
Line frequency:
50 Hz/60 Hz
Power consumption:
85 VA or less
Dimensions:
Approx. 212 (width)
88 (height)
400 (depth) mm
Mass:
5 kg or less
Safety:
Compliant with IEC61010-1 Ed.3
EMC:
Compliant with EN61326-1 classA
Vibration proof:
Compliant with IEC60068-2-6 2G
Option No.
Standard
OPT.32
OPT.42
OPT.44
Power voltage
100 V
120 V
220 V
240 V
Specify the option number when ordering.
Use a power cable and a fuse that are compliant with the safety
standard when changing the power supply voltage.