6.5. Discontinuing Measurement

If it is necessary to interrupt a blood pressure measurement for any reason (e.g
the patient feels unwell), press and hold the quick-release valve on the bulb at
any time. The device then immediately lowers the cuff pressure automatically.

6.6.  Irregular Heartbeat Detector

This function indicates an irregular heart beat. If the symbol of 
IHD          appears on the display that means that certain abnormalities in the
heartbeat frequency were detected during the measurement. In this case, the
result may deviate from your normal blood pressure – repeat the
measurement. In most cases, this is no cause for concern. However, if the
symbol          appears on a regular basis (e.g. several times a week with measurements taken
daily) we advise you to tell your doctor. Please show your doctor the following explanation:

Information for the doctor on frequent appearance of the Irregular Heartbeat Detector

This instrument is an oscillometric blood pressure monitor that also analyses pulse frequency during
measurement. The instrument is clinically tested. The symbol          is displayed after the measurement
if pulse irregularities occur during measurement. If the symbol appears more frequently (e.g. several
times per week on measurements performed daily) we recommend the patient to seek medical
advice. The instrument does not replace a cardiac examination, but serves to detect pulse
irregularities at an early stage.

7.  MEMORY FUNCTION

This semi-automatic blood pressure monitor automatically stores
the last result measured.

When the unit is off, press and hold the «

ON/OFF

» button and

the last measurement will be displayed.

8.  ERROR MESSAGES/TROUBLESHOOTING

If an error occurs during a measurement, the measurement is discontinued and a
corresponding error code is displayed (Example: Err 2).

Error No.

ERR 1

ERR 2

ERR 3

ERR 5

Possible Cause(s)/Solutions

The systolic pressure was determined, but then the pressure in the cuff went down to less than
20 mmHg (diastolic blood pressure can’t be measured). The tube may have loosened after the
systolic pressure has been determined. Further possible reason: pulse could not be detected.

Unnatural pressure impulses influence the measurement result. Reason: the arm was moved
during the measurement.

The difference between systolic and diastolic is excessive. Measure again carefully following
proper procedures. Consult your doctor if you still get unusual results of measurements.

The measuring signals are inaccurate and no result can be displayed. Read through the
checklist for performing reliable measurements and then repeat the measurement.

Summary of Contents for Advantage 6012N

Page 1: ...Directions for Use ADC 6012N AdvantageTM Semi Automatic Blood Pressure Monitor...

Page 2: ...lified to interpret blood pressure measurements This device is NOT intended to replace regular medical checkups It is recommended that your physician review your procedure for using this device Blood...

Page 3: ...Your level of blood pressure is determined in the circulatory center of the brain and adjusts to a variety of situations through feedback from the nervous system To adjust blood pressure the strength...

Page 4: ...as volatile diastolic pressure can still vary significantly In hypertensive individuals variations are even more pronounced Normally blood pressure is at its lowest during sleep and rises in the morni...

Page 5: ...ults of any one measurement Individual results may be influenced by spiking of your pressure due to diet anxiety or mis measurement resulting from excessive arm movement or misapplication of the cuff...

Page 6: ...ow at least 5 minute intervals between measurements Discard any reading that appears suspect and record the average of the remaining readings If this method is used be sure to note that the readings a...

Page 7: ...same time of day since blood pressure changes during the course of the day Measurements should be done after 5 minute rest to ensure accuracy 6 1 Fitting the Cuff Position the cuff flat on the table w...

Page 8: ...to a higher pressure see Inflating Further in section 6 4 When the device detects a pulse a heart symbol in the display starts to flash and a beep is heard for every heartbeat 6 3 Reading Measurement...

Page 9: ...ntly e g several times per week on measurements performed daily we recommend the patient to seek medical advice The instrument does not replace a cardiac examination but serves to detect pulse irregul...

Page 10: ...if the display is unusual then exchange them for new ones Check the connection of the cuff tube and connect properly 1 Fit the cuff correctly on the arm 2 Beforestartingmeasurementmakesurethatclothin...

Page 11: ...the manufacturer s warranty Note According to international standards your monitor should be checked for calibration every 2 years Contact ADC for an accuracy check 10 TECHNICAL SPECIFICATIONS Weight...

Page 12: ...is registered online at www adctoday com 12 QUALITY STANDARDS Device standard This device is manufactured to meet the European and United States standards for non invasive blood pressure monitors EN10...

Reviews: