If it is not possible to fit the cuff to your left arm, it can also be

placed on your right arm. However, all measurements should be made using
the same arm. Comparable blood pressure measurements always require the
same conditions (Relax for several minutes before taking a measurement).

6.2. Measurement Procedure:

After the cuff has been appropriately positioned and connected to the monitor,
the measurement can begin:

Press the «

ON/OFF

» button to turn the device on. You will see a flashing

1.

upward triangle 

.

You will hear 3 short beeps, and a steady “0” will be displayed.

2.

Grasp the bulb with your free hand and inflate the cuff to a

3.

pressure at least 30 mmHg higher than the expected systolic
pressure, during which an upward triangle will begin flashing.
If you do not know the value of this pressure, inflate to a value
between 160−180 mmHg.

Stop pumping when the upward triangle stops flashing. The

4.

device then starts to deflate.

Lay down the bulb and remain in a quiet and relaxed sitting position. The

5.

measurement now proceeds on its own and is taken automatically.

 If no measurement takes place and all that appears in the display is a

flashing, upwards pointing arrow, pump the bulb again and inflate to a higher
pressure (see “Inflating Further” in section 6.4). When the device detects a
pulse, a heart symbol in the display starts to flash and a beep is heard for
every heartbeat.

6.3. Reading Measurement Results:

When the measurement has been concluded, a long beep tone sounds. The
measured  systolic and diastolic blood pressure values, as well as the pulse are
now displayed. A downward triangle will flash, indicating the rest of air pressure in
the cuff can be released.

The cuff pressure can now be quickly released by pressing the
quick release valve on the bulb.

Note: 

To prolong the batteries’ life the device

switches off automatically if no button is pressed
for 3 minutes. Otherwise you can switch it off by
pressing the «

ON/OFF

» button.

6.4. Inflating Further:

If the cuff is not inflated sufficiently, the measurement is stopped after a few seconds and a
flashing arrow pointing upwards 

▲ 

appears in the display. It is then necessary to inflate at

least 20 mmHg higher than the previous value until cuff pressure displays again on the LCD.
The instruction to inflate further can appear several times if the pressure is still not sufficient.

Attention!

If the cuff is inflated too far (over 300 mmHg), HI appears in the display and a warning signal is
sounded. The cuff pressure must then be immediately released using the quick-release valve on
the inflation bulb!

Systolic
blood
pressure

Diastolic
Blood
Pressure

Pulse

Summary of Contents for Advantage 6012N

Page 1: ...Directions for Use ADC 6012N AdvantageTM Semi Automatic Blood Pressure Monitor...

Page 2: ...lified to interpret blood pressure measurements This device is NOT intended to replace regular medical checkups It is recommended that your physician review your procedure for using this device Blood...

Page 3: ...Your level of blood pressure is determined in the circulatory center of the brain and adjusts to a variety of situations through feedback from the nervous system To adjust blood pressure the strength...

Page 4: ...as volatile diastolic pressure can still vary significantly In hypertensive individuals variations are even more pronounced Normally blood pressure is at its lowest during sleep and rises in the morni...

Page 5: ...ults of any one measurement Individual results may be influenced by spiking of your pressure due to diet anxiety or mis measurement resulting from excessive arm movement or misapplication of the cuff...

Page 6: ...ow at least 5 minute intervals between measurements Discard any reading that appears suspect and record the average of the remaining readings If this method is used be sure to note that the readings a...

Page 7: ...same time of day since blood pressure changes during the course of the day Measurements should be done after 5 minute rest to ensure accuracy 6 1 Fitting the Cuff Position the cuff flat on the table w...

Page 8: ...to a higher pressure see Inflating Further in section 6 4 When the device detects a pulse a heart symbol in the display starts to flash and a beep is heard for every heartbeat 6 3 Reading Measurement...

Page 9: ...ntly e g several times per week on measurements performed daily we recommend the patient to seek medical advice The instrument does not replace a cardiac examination but serves to detect pulse irregul...

Page 10: ...if the display is unusual then exchange them for new ones Check the connection of the cuff tube and connect properly 1 Fit the cuff correctly on the arm 2 Beforestartingmeasurementmakesurethatclothin...

Page 11: ...the manufacturer s warranty Note According to international standards your monitor should be checked for calibration every 2 years Contact ADC for an accuracy check 10 TECHNICAL SPECIFICATIONS Weight...

Page 12: ...is registered online at www adctoday com 12 QUALITY STANDARDS Device standard This device is manufactured to meet the European and United States standards for non invasive blood pressure monitors EN10...

Reviews: