11.2 ZERO TRACKING
This parameter is used to set the range of the zero tracking. Zero tracking will
aid the scale to hold or return to zero and should be increased if large weights
are left on the scale or temperature is not consistent. The display will show “
ZEo
1
” to “
ZEo 8
”.
•
Press
[Tare/Zero]
to scroll through the settings. If it is set to “
ZEo 1
”
the zero tracking is at its smallest range and “
ZEo
8
” the highest.
•
Press
[Unit]
to confirm the selection and move to the next
parameter.
11.3 STABILIZATION RANGE
This parameter is used to set the range of the stability indicator. This is used to
determine when the scale will print automatically as well as indicate to the user
that the weight reading is stable. The display will show “
StA 1
” to “
StA 8
”.
•
Press
[Tare/Zero]
to scroll through the settings. If it is set to “
StA 8
”
then the stability is at its fastest and “
StA
1
” the slowest.
•
Press
[Unit]
to confirm the selection and move to the next
parameter.
11.4 STABILIZATION TRACKING
This parameter is used to set the size of the stability indicator tracking range.
This will aid the scale to remain stable once a result has been given. The
display will show “
Str 1
” to “
Str 5
”
•
Press
[Tare/Zero]
to scroll through the settings. If it is set to “
Str 1
”
then the stability range is at its smallest and “
Str
5
” the highest.
•
When the desired value is displayed, press the
[Print/Hold]
key to
select the value and exit the Technical Parameters. The display will
count down to zero and the scale will return to normal weighing.
•
If
[Unit]
is pressed instead of the
[Print/Hold]
key, the display will
ask for Pin for entering into the Factory Parameters. Switch off the
scale.