8.2 SETTING OF BACKLIGHT
•
Press
the
[Unit]
key to scroll to the second function “
bL x
” to set the
backlight function.
•
Press the
[Tare/Zero]
key to change the settings (“
x
” can be set to
1, 2 or 3) by scrolling through the options.
•
The user can select to have the backlight set to 1- off, 2 - on or 3 -
automatic.
When set to automatic, the backlight will be off unless a weight is
placed on the pan. When the weight is removed it will stay on for 10
seconds after the scale returns to zero. DEFAULT SET: 3 -
automatic
8.3 ENABLING OF UNITS
•
Press
the
[Unit]
key to scroll to the third function “
on Kg
” which is for
setting the units to be enabled and disabled.
•
Each weighing unit can be enabled or disabled so that the enabled
units can be selected during the operation of the scale by the user.
•
Use
[Tare/Zero]
to toggle between “
on
” and “
oFF
” and then the
[Unit]
key to confirm the selection and move to the next unit.
8.4 COMMUNICATION
ADDRESS
•
Press the
[Unit]
key after all units are set, to scroll to the fourth
function “
Add xx
” where “
xx
” is value for setting the ID for the RS-232
results output.
•
This function is used to set the communication address which is sent
via RS-232 as an ID code. There are 26 options to select from “
Add
1
” to “
Add 25
”. Set
“Add 0”
for no address. These relate to alphabet
letters for example 1=A, 2=B to 25=Y
•
Use
the
[Tare/Zero]
key to scroll through the options.
•
Press
[Unit]
to confirm the selection and move to the next function
for setting the Baud rate for sending the data via RS-232 interface.
14
© Adam Equipment Company 2006