Input commands format:
The scale can be controlled with the following commands. The commands must
be sent in upper case letters, i.e. “
T
” not “
t
”. Press the Enter key of the PC after
each command.
Z
Tares the scale to display the net weight. This is the same as
pressing
[Tare/Zero]
H
Sets the scale to hold if the hold function is enabled. Same as
pressing the
[Hold]
N
Sends the net weight to the RS-232 interface.
G
Sends the gross weight to the RS-232 interface.
T
Sends the tare weight to the RS-232 interface.
10.0 ERROR MESSAGES
During the initial power-on testing or during operation it is possible that the scale
may show an error message. The meaning of the error messages is described
below.
ERROR CODE DESCRIPTION
POSSIBLE CAUSES
П
П
П
П
П
П
A continuous beep is
heard.
Weight on the pan exceeds the
capacity of the scale. Remove the
weight from the pan.
CALEr
If the selected mass is
less than 10% or more
than 20% of the capacity
of the scale, an error
message “
CALEr
” will be
displayed and the scale
will return to zero.
Incorrect calibration mass. Repeat
the process correctly.
If an error message is shown, repeat the procedure that caused the message such
as turning the scale on, calibration or any other functions. If the error message is
still shown, contact your supplier for further support.