Summary of Contents for CPWplus 15

Page 1: ...Adam Equipment CPWplus SERIES P N 7949 Rev A5 June 2006 Software Revision 1 2 06 Adam Equipment Company 2006...

Page 2: ...Adam Equipment Company 2006...

Page 3: ...ECTING THE UNIT 11 6 6 PRINTING AND HOLDING FUNCTIONS 11 7 0 CALIBRATION 11 8 0 USER PARAMETERS 13 8 1 AUTO POWER OFF 13 8 2 SETTING OF BACKLIGHT 14 8 3 ENABLING OF UNITS 14 8 4 COMMUNICATION ADDRESS...

Page 4: ...2 Adam Equipment Company 2006...

Page 5: ...connected to the base The CPWplus platform scales are available in 6 models with different capacities All scales come complete with an easy to clean stainless steel top pan on a sturdy steel frame an...

Page 6: ...ipment or large motors Do not place near vibrating machinery Avoid high humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air mov...

Page 7: ...e them to attach the bracket to the base 3 Mount it on the wall Use two suitable screws not supplied to fix the bracket to the wall 3 4 SETTING UP THE PILLAR VERSION SCALE The CPWplus standard series...

Page 8: ...0 mA minimum or using batteries inside the scale unit 6 x AA size Remove the cover from the battery compartment under the scale Install six AA size batteries as shown below It is suggested that the al...

Page 9: ...ll range Units of Measure kg lb oz lb oz Stabilisation Time 2 seconds Operating Temperature 0 C to 40 C 32 F 104 F Humidity Up to 95 RH non condensing Power supply 6 x AA size non rechargeable batteri...

Page 10: ...ero Sets the display to true zero or net zero by storing the current weight in the tare memory 5 2 DISPLAY SYMBOLS Symbol Description Scale in auto zero range Stability indicator When shown the result...

Page 11: ...on is combined into one key Tare Zero You can press the Tare Zero key at any time to set a new zero point Re zeroing the scale is necessary only if small amounts of weight are still shown when the pla...

Page 12: ...tive value will be shown If the scale was tared just before removing the container this value is the gross weight of the container plus all products those were removed The zero indicator will also be...

Page 13: ...s simultaneously 7 0 CALIBRATION Occasionally the scale should be verified that it is weighing correctly by measuring a known mass of approximately the scale capacity Tare the scale place the mass on...

Page 14: ...The digit stops flashing If the selected mass is less than 10 of the capacity of the scale an error message CALEr will be displayed and the scale will return to zero Repeat the process correctly Plac...

Page 15: ...back to on again or press the Print Hold key 8 1 AUTO POWER OFF The first function is to set the auto power off function The display will show Pr oFF or Pr on DEFAULT SET Pr oFF Press Tare Zero to to...

Page 16: ...be enabled or disabled so that the enabled units can be selected during the operation of the scale by the user Use Tare Zero to toggle between on and oFF and then the Unit key to confirm the selection...

Page 17: ...or sending data to RS 232 interface 8 6 SELECTION OF BIT RATE AND PARITY Press the Unit key to scroll to the sixth function Par x to select the bit rate and parity used for sending data to RS 232 inte...

Page 18: ...o the next function for setting the Hold function 8 8 SELECTION OF HOLD FUNCTION Press the Unit key to scroll to the eighth function Hod x to set the Hold function Hod x appears on the screen See the...

Page 19: ...ight is stable Hold is released if Print Hold is pressed or time expires as per Hti setting Print continuously Hold occurs automatically when the weight is stable Hold is released if Print Hold is pre...

Page 20: ...e socket Pin 3 Output Pin 2 Input Pin 5 Signal Ground Normal Output add A G W 2 000 kg G W is Gross Weight lf Includes 3 line feeds lf lf add A N W 1 000 kg N W is Net weight lf Includes 3 line feeds...

Page 21: ...ting or during operation it is possible that the scale may show an error message The meaning of the error messages is described below ERROR CODE DESCRIPTION POSSIBLE CAUSES A continuous beep is heard...

Page 22: ...return to normal weighing turn the scale off and back to on again or press the Print Hold key 11 1 FILTER This parameter is used to set the speed of the display filtering For poor environments the fil...

Page 23: ...1 to StA 8 Press Tare Zero to scroll through the settings If it is set to StA 8 then the stability is at its fastest and StA 1 the slowest Press Unit to confirm the selection and move to the next para...

Page 24: ...ssed by a qualified technician To exit the Pin Code entry when the display shows Pi the scale must be switched off 13 0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and acces...

Page 25: ...ased This part of information should always be available for any future correspondence We suggest you to fill in this form as soon as the unit is received and keep a print out in your record for ready...

Page 26: ...th correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse a...

Page 27: ...ovide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not inst...

Page 28: ...eb aeadam de Adam Equipment S A Pty Ltd 7 Megawatt Road Spartan EXT 22 Kempton Park Johannesburg Republic of South Africa Phone 27 0 11 974 9745 Fax 27 0 11 392 2587 e mail sales adamequipment co za A...