6.1.4
F10
‐ Factory Calibration Value
This calibration is stored in a separate location from
the user calibration (F4).
If a scale has been disrupted it is possible to recall this calibration in order to
have a scale that has the original calibration restored.
This calibration can be done at any time without checking for being within
10% of the previous factory calibrations. This value will also overwrite the
user calibration value as set in F3. When the user calibration is performed
the results must agree within 10% of the factory calibration. This prevents
calibration with no weight on the scale or with grossly wrong weights.
When the display is showing:
“
F10
” “
FACt
”
“
Set
”
Press the
[Total]
key. The display will show:
“
0.000
” “ ” “
xxxxxx
”
Make sure no other weights are on the scale.
Press the
[Total]
key.
The displays will show the calibration weight required, equal to the
full scale capacity.
“
20.000
” “ ” “
xxxxxx
”