4.5 For best accuracy follow these suggestions.
•
Samples should be tested more than once and the average value taken.
•
Only distilled water should be used.
•
The water should be at a stable temperature.
•
The sample must be degreased.
•
The sample should not be porous. If it is porous then special preparation might be
necessary.
•
Be certain the sample is completely covered by about 10mm of the liquid.
•
Be certain any air bubbles are removed.
•
Dry the sample completely between tests.
•
Be very careful not to transfer water to the top pan when using tweezers.
•
Samples should be 5 grams or greater.
•
Degrease the glass float before using.
•
Dry the float completely between each test. Be careful not to put a static electric charge
on the glass float. It may be necessary to use an anti-static cleaner on the float.
•
Wash the beaker before filling it with any liquid.
•
Place the samples on the weighing pans carefully to avoid shocks to the balance.