Doc : AM210-001E Release: 1.4 updated : 2015
Proceed in the same way as routine thickness measurement:
-
Open the DR10 test compartment cover
-
Identify the location of the tube serial number
-
Carefully place
brand new heater tube without deposit
ion the support, the tube serial number
must be located on the right.
-
Close the test compartment door and press OK.
-
Wait till the tube identification screen appears.
-
Use the scroll buttons to rotate the tube and see the tube serial number in the dedicated
window.
-
Type in the tube serial number corresponding to the number you read and press ENTER
-
The automatic baseline calibration procedure will be performed
-
When completed, press Ok
CAUTION!
The test compartment door must be closed during all calibration process.
-
Remove the brand new heater tube from the DR10 test compartment and place it in its original
container.
Calibration information will be automatically stored in the instrument memory.