Doc : AM210-001E Release: 1.4 updated : 2015
Set pre-programmed lists
The “
List Settings
” button allows entering information which future will be used in drop-down lists.
There are five sections in this menu:
Operators
- identification of the person performing the test. Maximum 15 characters.
Sample Type
- identification of jet fuel type for reporting. Maximum 15 characters.
Temperature
- the test temperature (in °C) of jet fuel oxidation in thermal oxidation stability rig during
D3421 test. Maximum 3 digits.
Test Rig ID -
identification of the thermal oxidation stability rig. Maximum 15 characters.
Touch the entry area and the virtual keyboard will appear.
Type-in the data for each field and press ENTER.
Press Add button to add any data in the list.
If you want to delete a data already entered, select this data and press Del button.
In order to memorize the modifications,
press SAVE button
prior leaving the menu.
User Defined Steps Numbers –
Please see the “
TEST OPERATION with User Defined Matrix”
Section.
Set password
The access to specificDR10menus is password protected in order to prevent unauthorized changes of
critical instrument settings.
The password is a 5 digits code.
The factory set password is 00000.
This password is active when you receive the instrument.
In order to change the password, go to the menu SETUP/PASSWORD SETTINGS.