2.
Pass Count:
The number of times to ignore triggering signals that match the
trigger parameters. It is set to 0 by default to indicate that it is not ignored.
Multi-Level Trigger
Multi-level trigger is composed of multiple single level trigger. This function has maximum
16 levels, each level must be set separately and set in the same way as the single level
trigger. When adding a new level, press the button on the top to select the relationship
between each level. The relationship between each state can be a continuous trigger
(Next IF) or a non-continuous trigger (Then IF).
1. Schematic diagram of the current set of trigger conditions
2. Trigger conditions setting
As shown in the picture below, the first and the second classes are continuous
trigger, the relationship between the second and third classes are
non-continuous trigger, and the third, fourth, fifth and sixth classes are
continuous trigger.
Difference between continuous trigger and non-continuous trigger:
Continuous trigger: The signal captured by two adjacent sample clocks must meet
the conditions to trigger.