10
AR100 Operation Manual
3.4.1
Serial Communications Check
If no information is received over the serial port, check the power supply
and serial wire connections. The sensor may be in a configuration that
prevents serial communication, such as being set at the wrong baud rate
or is in a polling mode.
Use the Demo and Configuration Software to restore the unit to factory
defaults. Otherwise, the
04h
command will restore factory defaults to the
device.
3.4.2
Sensor Output Check
If the sensor output value is in error, check that the sensor and target are
stationary and stable and that the laser beam is hitting the target.
The sensor may need to warm up for 5-10 minutes
before reaching
full accuracy. Leave it on for a few minutes and re-check the sensor
accuracy.