
AR4000 User’s Manual
23
LLL004001 – Rev 2.7
If both the serial and current loop outputs are enabled, the maximum output rate
is 1500 microseconds per sample for calibrated output.
For uncalibrated output, the maximum serial output rate is 400 microseconds per
sample, and the maximum current loop update rate is 300 microseconds per
sample.
11.2 Resolution
Range resolution is limited by the sample rate selected and the maximum range
to be measured. Specifying faster sample rates or longer maximum ranges will
result in poorer resolution. Calibrated range output is always transmitted in units
of .01 inches. However, resolution will show up as “steps” in the output values
transmitted by the sensor. For the current loop output, resolution may also be
limited by the resolution of the digital to analog converter used. This may
become noticeable for large span settings,
since the inherent sensor resolution for
low sample rates is better than 1 part in 10,000. The table below shows the
resolution that will be obtained for a variety of sample rate and maximum
distance settings.
Current loop output is generated with a resolution of 1 part in 4000, and is linear
with respect to measured range to 1 part in 1000. Selectable zero and span allow
full resolution over any distance span.
Sensor amplitude, ambient light, and temperature are 10 bit samples and are
internally updated at the same frequency as the range measurement up to 10,000
samples per second.
Maximum Attainable Sample Rates, samples/second
Resolution, Range
inches
6 Feet
30 Feet
55 Feet
.0062
2304
677
390
.0125
4609
1355
781
.0250
9218
2711
1562
.0500
18346
5422
3125
.1000
36873
10845
6250
.2000
50000
21691
12500
.4000
50000
43382
25000
.8000
50000
50000
50000