7
Battery Charge Indicator:
The Battery Charge Indicator consists of
four LEDs labeled “25%”, “50%”, “75%”, and “100%”.
NOTE:
Charge level indication is approximate.
Hour Meter Button:
The POC will read and display the number
of hours that the unit has been used. (Refer to Routine Maintenance
section.)
Nasal Cannula:
A standard single lumen nasal cannula or equivalent must
be used with the POC to provide oxygen from the concentrator. A standard
single lumen cannula, 7-foot (2-meter), is recommended for use.
†
For
a replacement cannula, please contact your local medical equipment
supplier. Follow cleaning and care instructions provided with the nasal
cannula.
CAUTION:
Use of some administration accessories not specified for use
with this oxygen concentrator may impair the performance.
* Reference Alarm Indicators Table for explanation of low battery alarms.
†
Nasal cannula may not be included. Based on international requirements.
Battery Check LED
Charge Level
25% LED
50% LED
75% LED
100% LED
Off
Off
Off
Off
0%*
Blinking
Off
Off
Off
1% -12%*
On
Off
Off
Off
13% - 25%
On
Blinking
Off
Off
26% - 37%
On
On
Off
Off
38% - 50%
On
On
Blinking
Off
51% - 62%
On
On
On
Off
63% - 75%
On
On
On
Blinking
76% - 97%
On
On
On
On
98% - 100%