11
3. It can take a fully discharged
Internal Battery up to 4.5 hours
to reach a full charge. Press the
Battery Check button to check the
charging process.
4. When the Internal Battery is fully
charged, the INT Battery Check LED
will turn off.
5. It is recommended to recharge
the Internal Battery, even if only
partially depleted, as often as possible. The Internal Battery
cannot be overcharged, so it is okay to leave it plugged in via
an AC or DC Adapter/Charger continuously.
EXTERNAL BATTERY CHARGING
1. Connect the External Battery to the External Battery Port on the
POC. Then connect the unit to either an AC or DC Adapter and
corresponding charger to a suitable outlet. To prolong battery life
and maintain communication with the POC, it is recommended to
charge the External Battery when you charge the POC’s Internal
Battery.
2. Verify that the Battery Check LED labeled EXT is solid green and the
Power LED is flashing green. This indicates that an external power
source is recognized and the battery is charging.
NOTE:
If the EXT LED does not illuminate after the INT LED turns off,
the external battery isn’t communicating with the POC. To enable
communications and start charging
the external battery: a) Power off the
POC. b) Press and hold the Battery
Check button and verify the INT LED
illuminates when depressed. c) Using
two fingers, depress the “Mute” and
LPM1 buttons simultaneously for
1 to 2 seconds, then release. EXT
LED should beginning blinking and
will change to solid Green when
connection is made. The solid green
LED on INT indicates the external battery is charging. If not successful,
try again. If this procedure does not work for you, call Inova Labs
Customer Care at 512-617-1665 for assistance.
OPERATING
INSTRUCTIONS
continued