25
Guidance and
manufacturer’s
declaration
–
electromagnetic immunity Table #4
The Action Trackchair and Trackstander is intended for use in the electromagnetic environment
specified below. The customer or the user of the Trackchair/Trackstander should assure that it is used
in such an environment.
IMMUNITY test
IEC 60601 TEST
LEVEL
Compliance
level
Electromagnetic
environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the
Trackchair/Trackstander,
including cables, than the
recommended
separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended separation
distance
d =
1,2
P
d =
1,2
P
80 MHz to 800 MHz
d =
2,3
P
800 MHz to 2,5 GHz
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
a.
should be
less than the compliance
level in each frequency
range.
b.
Interference may occur in
the vicinity of equipment
marked with the following
symbol: