Overall dimensions:
Control unit:
Width: 270mm
Height: 100mm
Height with tanks: 131mm
Depth: 246mm
Weight: 3Kg
Footswitch:
Width: 173mm
Height: 46mm
Depth: 177mm
Cord length: 2,000mm
Weight: 610g
Handpiece cord: 2,000mm
Tank capacity: 300ml each
Temperatures:
Operating: +10°C to +40°C.
Storage: -20°C to +70°C.
Humidity:
Operating: 30% to 75%.
Storage: 10% to 100% including condensation.
IV - INSTALLATION / FIRST
USE
4. 1 UNPACKING THE DEVICE
Upon reception of the device, check for any
damage caused in transit.
Contact your supplier if necessary.
4. 2 RECOMMENDATIONS
Have your connections to the electrical and air
supplies made by a certified dental installation
technician.
The electrical connection of the P M
AX
N
EWTRON
XS
must comply with applicable standards. A
grounded power supply socket must be used.
The device must be connected to the dental
office's medical-quality air distribution supply.
4. 3 INSTALLATION
Important:
Never rotate the handpiece connector on its
cord or you may damage the device.
Do not place the P M
AX
N
EWTRON
XS close to or
on top of another device.
Do not intertwine the power cord with the
footswitch cord or the handpiece cord.
- Place the control unit in position on a fixed,
horizontal surface or with a slope of no more
than 5°.
- Check that the power switch is in position O
(off).
- Connect the power cord to the receptacle of
the device.
- Connect the power cord to a socket with a
ground pin.
- Connect the multifunction footswitch to the
footswitch connector. Fully tighten the screws.
- Position the footswitch so that it is easily
accessible for your feet.
- Check that there is no humidity in the
handpiece cord connector connections. If there
is, dry it out by wiping, then blowing with the
multifunction syringe.
- Connect the handpiece cord to the connector
on the device located on the front panel,
making sure that the two marks on the cord
plug and the connector symbol are lined up.
9