Product description

2.1 

Product code

Hydrocont HP4 [01][02][03][04][05][06][07][08][09][10][11][12][13][14][15][80][94/95/96/97/98/99]

01

Application type

S

Standard

02

Sensor / material diaphragm

C

capacitive – front-flush / ceramic Al2O3 96%/99,7%

03

Approval

S

Standard

04

Process connection

0

without

05

Material process seal

1

FKM/FPM

05

3

EPDM, FDA listed

06

Material process connection

V

CrNi-steel

06

D

CrNi-steel, duplex, sea water resistant

07

Terminal enclosure

0

without

07

1

Field enclosure 3pole, IP65, PS [09-A] +[10-0]

07

2

Field enclosure 6pole, IP65, PS [09-V], [09-A]+[10-1]

08

Measuring range pressure

01

0…100 mbar

08

02

0…200 mbar

08

03

0…400 mbar

08

04

0…600 mbar

08

05

0…1 bar

08

07

0…2 bar

08

08

0…4 bar

08

09

0…6 bar

08

10

0…10 bar

08

12

0…20 bar

08

0A

0…1 mwc

08

0B

0…2 mwc

08

0C

0…4 mwc

08

0M

0…5 mwc

08

0D

0…6 mwc

08

0E

0…10 mwc

08

0F

0…20 mwc

08

0L

0…25 mwc

08

0G

0…40 mwc

08

0J

0…50 mwc

08

0K

0…60 mwc

08

0H

0…100 mwc

09

Electronic – Output

A

Current 4…20mA, FSK, 2-wire, over voltage protection

09

V

RS485 Modbus-RTU, 4- wire, over voltage protection

10

Electronic – Function

0

without

10

1

Temperature Pt100 class B, 3-wire – IEC 60751 [09-A]

10

3

Temperature ≤ ±0,5K [09-V]

11

Process temperature

0

Standard -20°C...+70°C (-4°F… 158°F)

12

Pressure type

R

Gauge pressure

13

Measurement pressure - accuracy

1

0,2%

13

3

0,1%, linearization protocol

13

6

Xcellence – 0,05% ([08] ≥ 200mbar/2mWS), linearization protocol

14

Electrical connection

K

Cable, confection stranded wires (length L1 +240mm)

14

H

Cable, confection Hydrolog HLF4 (length L1  -360mm)

14

0

Cable, without confection (incl. confection kit)

15

Material extension cable

A

Cable sheath PE

15

B

Cable sheath PUR

15

G

Cable sheath PUR, increased diffusion-tightness

80

Length L1

-###.###

mm (≤ 300.000mm)

94

 Additional option

-SF

LABS-free, silicone-free / paint compatible version

95

 Additional option

-ML

Measurement point designation / TAG – Laser marking

96

 Additional option

-MZ

Material test certificate – EN10204 3.1 

97

 Additional option

-WT

Factory certification – drink water suitability

98

 Additional option

-KF

Configuration / Preset

99

 Additional option

-WK

Factory calibration – calibration certificate

Summary of Contents for Hydrocont HP4SC

Page 1: ...OPERATING MANUAL BA03 22 Hydrocont HP4SC Hydrostatic filling level transmitter with capacitive ceramic measuring membrane...

Page 2: ...6 3 4 Field enclosure 7 4 Electrical connection 7 4 1 Electronic output 09 A 4 20mA FSK interface 7 4 2 Electronic output 09 V Interface RS485 Modbus RTU 8 5 Operation 9 5 1 Electronic output 09 A 4 2...

Page 3: ...ntal contamination can result Also the characteristics of the instrument can be impaired An inappropriately use disregarding the Operating manual and the technical rules using under qualified personne...

Page 4: ...08 0H 0 100 mwc 09 Electronic Output A Current 4 20mA FSK 2 wire over voltage protection 09 V RS485 Modbus RTU 4 wire over voltage protection 10 Electronic Function 0 without 10 1 Temperature Pt100 c...

Page 5: ...ead out at the version 09 V RS485 Modbus RTU by the digital interface or the resistance signal can be evaluated at the version 09 A current 4 20mA FSK in parallel to the pressure conditioned analogue...

Page 6: ...perature calmed places High process temperature steps can produce short time higher measuring signal deviations At underrun of the dew point e g cold process medium at high environmental temperature t...

Page 7: ...r to receive two or more digital actualizations per seconds from an intelligent field device The digital signal does not disturb the 4 20mA The communication protocol offers two simultaneous communica...

Page 8: ...the data wires The optimum resistor value depends on the wave impedance of the used cable but a value of 120 Ohm is a popular choice The polarisation network is necessary to ensure suitable potential...

Page 9: ...at the output by 100 Start value LRV Nominal lower adjust value LRV URV 4mA Span 25 End value URV Nominal upper adjust value LRV URV 20mA Span 25 5 2 Electronic output 09 V Interface RS485 Modbus RTU...

Page 10: ...nd device in for repair Supply voltage Operating voltage available Switch on resp repair operating voltage Check terminals resp repair Operating voltage too low too high Adapt operating voltage Load r...

Page 11: ...ility negligible Influence of auxiliary power 0 002 FSI V Influence of temperature Tk zero 0 015 FSI K 0 75 FSI Tk span 0 015 FSI K 0 5 FSI 0 4bar 4mwc 0 8 FSI 0 4bar 4mwc Influence of mounting positi...

Page 12: ...o PE 500VDC on request Rated leakage current 10kA 8 20 s Protection class III Pollution degree 4 Altitude 5000m above NN MTTF 463 4 years Weight 0 3kg L1 x 0 068kg m Materials Process wetted Ceramic A...

Page 13: ...rolsystem de 49 0 8721 9668 0 ACS CONTROL SYSTEM GmbH Lauterbachstr 57 84307 Eggenfelden Germany 8 Dimensions 8 1 Sensor Electrical connection 14 K Confection stranded wires Electrical connection 14 H...

Page 14: ...rNi steel Art No 65001899 Dimensions in mm 9 2 Mounting screw VSM 1000 Thread ISO 228 1 G1 CrNi steel Gasket FKM FPM IP65 IP67 Art No 611000053 VSM 1500 Thread ISO 228 1 G1 CrNi steel Gasket FKM FPM I...

Page 15: ...rail width 22 5mm Art No 171000004 WTAU 100 U0 Pt100 isolation amplifier active 20 253Vuc In Pt100 Out 0 4 20mA 0 10V adjustable switch cabinet assembly at standard mounting rail width 22 5mm Art No 1...

Page 16: ...N E R F O R M E A S U R E M E N T A U TO M AT I O N ACS CONTROL SYSTEM GmbH Lauterbachstr 57 84307 Eggenfelden Germany info acs controlsystem de www acs controlsystem com 49 0 8721 9668 0 FEEL FREE T...

Reviews: