Service
6.1
Maintenance
At appropriate use, the device is free of maintenance.
Solid coatings on the diaphragm can lead to faulty measurement results. In this case the diaphragm must be regularly
cleaned. Don’t use sharp resp. hard tools, pressured air or aggressive chemicals. For dismounting the device see chapter
“Dismounting”.
6.2
Dismounting
Use suitable protective clothing, e.g. goggles, gloves.
WARNING:
Dismount the device only when power supply is off.
WARNING:
Let the device and the system cool down sufficiently fore dismounting it. There is a risk of hot surfaces as well as
dangerous and hot media escaping.
WARNING:
Dismount the device only when the system is pressureless. There is a risk of fast escaping media resp. pressure
blow.
After dismounting the diaphragm must be fitted with protective caps.
6.3
Troubleshooting / Repair
The operator of the system is responsible for taking suitable measures to rectify faults.
In case of malfunction check:
Component / area
Check
Rectification
Enclosure
Damage
Replace device resp. send in for repair
Pressure diaphragm
Pollution
Clean device resp. send in for repair
Damage
Replace device resp. send in for repair
Process seal
Damage
Replace process seal
Use other seal material if necessary
Compensation capillary
Pollution
Clean capillary resp. send device in for repair
Supply voltage
Operating voltage available
Switch-on resp. repair operating voltage
Check terminals resp. repair
Operating voltage too low / too high
Adapt operating voltage
Load resistance too high
Reduce resistance
Increase operating voltage
Connection cable damaged
Send in device for repair
For dismounting the device see chapter “Dismounting”.
If the malfunction cannot be eliminated, please contact the manufacturer.
A repair may only be carried out by the manufacturer.
6.4
Return
Enclose necessary information’s for return:
•
An exact description of the application.
•
The chemical and physical characteristics of the product.
•
A short description of the occurred error.
Before returning the device, the following measures must be performed:
•
Adhesive product residues e.g. caustic, toxic, radioactive etc. must be removed.
•
A returning must be refrained, if it is not possible by 100% to remove the unhealthily product completely.
•
The device must be packed damage-proof.
6.5
Disposal
This instrument is not subject to the WEEE directive and the respective national laws. Hence, pass the instrument directly on
to a specialized recycling company and do not use the municipal collecting points.
1100
2 / UInt16
Status
Bit 0: 1 = Invalid Measure Value PV
Bit 1: 1 = Invalid Measure Value SV
1101
2 / UInt16
Unit
PV = Pressure
1102
4 / Float
Measure Value
PV = Pressure
1104
2 / UInt16
Unit
SV = Temperature
1105
4 / Float
Measure Value
SV = Temperature
Holding Register – Values
2 Byte – Uint16 – R/W
Address
Description
Default
Comment
2020
Damping Tau
1000
ms / PV = Pressure
2021
Damping Tau
1000
ms / SV = Temperature