Technical manual   

BA 1118 

Filling level 

SAT 

Electrode probe 

 

for conductive limit level detection 
in electrical conductive liquids

 

Detects up to six limit values simultaneously 

 

Useable 

 

for leakage or overflow protection in container 

 

for minimum, maximum resp. multilevel detection in container 

 

as pump protection, resp. dry run protection in pipelines 

 

for two-position-control of pumps 

 

Wide application range 

 

for conductivities higher than 1 µS/cm  

 

for process temperatures from -15 °C to +150 °C 

 

for process pressures from -1 bar to +10 bar 

 

materials also for aggressive filling material 

 

ATEX II 1 G Ex ia IIB/IIC T6

…T1 Ga 

Certification for the use in explosion hazardous areas 

 

Integrated wire break monitoring 

L a u t e r b a c h s t r .   5 7  

–   8 4 3 0 7   E g g e n f e l d e n   –   G e r m a n y  

T e l :   + 4 9   8 7 2 1 / 9 6 6 8 - 0  

–   F a x :   + 4 9   8 7 2 1 / 9 6 6 8 - 3 0  

i n f o @ a c s - c o n t r o l s y s t e m . d e

 

 

w w w . a c s - c o n t r o l s y s t e m . d e

 

 

Summary of Contents for SAT

Page 1: ...position control of pumps Wide application range for conductivities higher than 1 S cm for process temperatures from 15 C to 150 C for process pressures from 1 bar to 10 bar materials also for aggres...

Page 2: ...SAT page 2 of 9 Index Application field 3 Function 3 Safety notes 3 Installation 4 Maintenance 4 Repair 5 Electrical connection 5...

Page 3: ...trode probe is mounted directly in the wall of the container or of the pipe by using the respective process connection or installed over the filling material by using a suitable mount The alternating...

Page 4: ...ce is installed and operated in explosive hazardous areas the general Ex construction standards EN IEC 60079 14 VDE 0165 these safety notes and the enclosed EC conformity certificate must be observed...

Page 5: ...y if the filling material is strongly fluctuating Use for the stabilization suitable isolating spacers The distance between the spaces should be not more than 0 5 m At horizontal side mounting into a...

Page 6: ...parated from power leading wires At strong electromagnetic irradiation use principally a cable with shield Connect the cable shield only at one side to earth A suitable evaluation device must by conne...

Page 7: ...5 C Limitation by material Environmental temperature range Terminal enclosure PP 5 100 C Process temperature maximum 15 C 150 C observe limitations Limitation by variant Process temperature range ATEX...

Page 8: ...rocess connection G15 G 1 Process connection G20 G 2 Process connection G10 G 1 Process connection G12 G Terminal enclosure 60mm only with material POM Terminal enclosure 40mm with terminal enclosure...

Page 9: ...others on request Material process connection terminal enclosure D POM 40mm for G G 1 resp 80mm for G 1 G 2 E POM 60mm for G G 1 P PP 40mm for process connection G G 1 M PP 80mm for process connection...

Reviews: