"no" by Up and Down keys, and the flashing indicates the selection. Press the "Enter" key to save
the parameters or not, exit the programming mode and enter the operation mode.
6.4.3 AID10 alarm and display key operation
(1) After AID10 and AIM-M10 are connected through RS485 communication, the On light flashes,
indicating that the communication is normal. If the operation lamp is normally on, it means the
communication is abnormal.
(2) When AID10 detects the fault mark issued by AIM-M10, the corresponding indicator light flashes,
and the buzzer rings. Press the Mute key to turn off the buzzer.
(3) When the system is running normally, press the Test key to start the AIM-M10 insulation monitor
self-test.
6.4.4 AID150 alarm and display key operation
1) Description of the Operating Interface
After the system is powered on, if there is no fault alarm, AID150 shows the normal operation
interface as shown in the following figure. The black boxes in the figure indicate that the
corresponding address serial number is connected to the instrument communication, and the black
boxes indicate that there is no instrument connection, or that the communication is not connected.
When the insulation monitor or residual current monitor detects the fault, AID150 displays the
corresponding alarm interface and sends out the corresponding sound and light alarm.
System normal
2015-07-02 12:30:45
--------------------
--
--
System fault(01/02)
Fault type:Insu
BRK OL OT
Loc.:ICU Bed:04
System fault(02/02)
L1:OK
L2:OK
L3:OK
L4:ORC
Loc.:OR Room:06
Normal system fault indication(AIM-M10) fault indication(AIM-R100)