parameters
Fuse wire
80A
63A
50A
35A
25A
Primary winding
resistance
<55mΩ
<64mΩ
<80mΩ
<131 mΩ
<245mΩ
Secondary
winding
resistance
<45mΩ
<64mΩ
<80mΩ
<116 mΩ
<228mΩ
Iron loss
<150W
<105W
<107W
<77W
<55W
Copper loss
<230W
<200W
<170W
<125W
<120W
Efficiency
>96%
>96%
>96%
>96%
>95%
Maximum ambient
temperature
<40℃
<40℃
<40℃
<40℃
<40℃
No-load
temperature
rise
<36℃
<33℃
<31℃
<26℃
<22℃
Full load
temperature
rise
<65℃
<76℃
<67℃
<62℃
<55℃
Noise grade
<40dB
<40dB
<40dB
<40dB
<40dB
4.2 Technical parameters of AIM-M10 medical insulation monitor
Refer to Table 4.
Table 4 Technical parameters of AIM-M10 insulation monitoring instrument
Auxiliary
power supply
Voltage
AC220V
(±10%)
Temperature
measurement
Thermistor
PT100
Frequency
50/60Hz
Measuring range
-50—+200
℃
Power
consumption
<5W
Alarm value range
0—+200
℃
Insulation
monitoring
Measuring
range of
insulation
resistance
10-999kΩ
Alarm output
Output mode
1 relay output
Absolute
percentage
error
0—±10%
Contact capacity
AC 250V/3A DC
30V/3A
Alarm value
50—999kΩ Environment
Operating
-10—+55
℃