OMNISWD
®
USER MANUAL
21
COPYRIGHT © 2019-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Safety Features / Error Messages
The OmniSWD
®
is microprocessor controlled; it continually “self-checks” the operation of the system. If an error in
the function of the microprocessor is detected the shortwave power will be immediately turned off and the word FAIL
will be shown on the display. If the unit repeatedly fails after cycling the power, then qualified service personnel should
be called. The unit incorporates a safety mechanism in case of overheating. If the unit overheats the treatment is
automatically stopped, the buzzer sounds intermittently, and “HOT” is shown on the display.
Coaxial Cable - Applicator Head
Error Warning
This pop-up window appears if a treatment is attempted
without the applicator being plugged in (or if the
cable/applicator is damaged). It will also appear if a fault
occurs in the cable or applicator during a treatment.
Treatment will be prevented from starting unless a
functioning applicator/cable combination is plugged in.
If it persists, then contact ACP Customer Support at
(1-800-350-1100).
System Fault Warning
Service Due Pop-Up
This is just a message that appear as a flag from
the hours of use. The message is not an indication
that something is wrong with your operating
system or device. For service please contact
ACP Customer Service at 1-800-350-1100 ext.3.
Process to Reset the Service Due Timer:
1. Start the system
2. Go to settings
3. Go to maintenance
4. Insert the password: 12341924 and save
5. Go to service date
6.
Press reset hours counter and confirm
This pop-up window appears if the machine detects a
communications error between its two processors. This
should be a very rare event if it happens at all, and may
not recur if the mains power is switched off and on
again. If it persists, then contact ACP Customer
Support at (1-800-350-1100).