43
Display
Causes
Solution
Temperature has
exceeded the
operating temperature
of the system
Move to a cooler environment.
Repeat the test.
Temperature is below
the operating tempera-
ture of the system
Move to a warmer environment.
Repeat the test.
Battery is low but has
enough power to run
20 more tests
Test results will still be accurate.
But replace the battery as soon
as possible.
Battery is discharged
and the meter does not
allow more tests until
replacement with a
new battery
Replace the battery. Repeat the test.
Meter electronics
failure
If the problem persists, please
contact Technical Support at
1-800-838-9502.
Non
On Call
®
Pro
test
strip inserted in the
meter
Please make sure you use the
On Call
®
Pro
test strip with the
On Call
®
Pro
Blood Glucose Meter.
If the problem persists, please
contact Technical Support at
1-800-838-9502.
Test strip damaged or
calibration error
Please test again by using a new
strip. If the problem persists, please
contact Technical Support at
1-800-838-9502.
Communications
failure
There is an error in transferring
data to the PC. See the package
insert included in the
On Call
®
Diabetes Management Software
for troubleshooting.
Meter data port is
plugged in with
the data transfer
cable when meter is
in waiting for sample
application mode with
strip already inserted
into the meter strip port
Unplug the data transfer cable from
the meter’s data port. Then remove
the strip. Reinsert the strip into the
strip port for testing. If the problem
persists, please contact Technical
Support at 1-800-838-9502.