Hypo warning; Test Reminder 1, Test Reminder 2, Test Reminder 3, Test Re-
minder 4, Test Reminder 5.
If any of the above signs of physical or functional deterioration are observed on
the meter after performing the cleaning and disinfection procedure, please stop
using the meter and contact Technical Support at 1-800-838-9502
for assistance.
If the meter displays an error message, if control solution test results fall
outside of the indicated control range or if you suspect that the test
results are inaccurate after performing cleaning and disinfection, please stop
using the meter and contact Technical Support
at 1-800-838-9502, 24 hours a day, 365 days a year.
Note:
All parts of the kit are considered bio-hazardous and can potentially
transmit infectious diseases, even after you have performed cleaning and
disinfection. Please follow your institution’s procedures and precautions
when handling the meter.
Important:
After disinfection, gloves should be removed and hands should
be thoroughly washed with soap and water before proceeding to the
next patient.
For more information, refer to “FDA Public Health Notification: Use
of Fingerstick Devices on More than One Person Poses Risk for
Transmitting Bloodborne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevices/ Safety/AlertsandNotices/ucm224025.htm
and “CDC Clinical Reminder: Use of Fingerstick Devices on More than One
Person Poses Risk for Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html.
General Maintenance
This Blood Glucose Meter is intended for multiple patient use by health care
professionals in health care facilities. Always remember to pre-clean and
disinfect the meter after each use. This is important to potentially prevent the
transmission of infectious diseases.
Avoid getting liquids, dirt, and blood or control solution into the meter’s strip
or data ports.
The
On Call
®
Pro
Blood Glucose Meter is a precision electronic instrument.
Please handle it with care.
Safety Lancets
Please use single-use, auto-disabling safety lancets for blood sampling.
There is no maintenance required for single-use, auto-disabling
safety lancets.