17
9.
Hyper Indicator:
The meter comes with the Hyper indicator feature dis-
abled. Press the left (◄) button to turn the Hyper indicator “On” or “Off”.
Press the right (►) button to confirm your selection. When the Hyper
indicator is “Off”, pressing the right (►) button will go to the next Hypo
indicator set up. When the Hyper indicator is “On”, pressing the right (►)
button will go to the Hyper indicator level set up. At the Hyper level set
up, press the left (◄) button to adjust the Hyper level. Press the right (►)
button to go to the Hypo indicator set up.
Note:
The meter allows the hyperglycemia level to be set as low as 120
mg/dL. The hyperglycemia level should be above the hypoglycemia level.
The proper hyperglycemia blood glucose level for each patient should be
determined by their physician.
10. Hypo Indicator: The meter comes with the Hypo indicator feature dis-
abled. Press the left (◄) button to turn the Hypo indicator “On” or “Off”.
Press the right (►) button to confirm your selection. When the Hypo in
-
dicator is “Off,” pressing the right (►) button will go to the Test Reminder
set up. When the Hypo indicator is “On”, pressing the right (►) button will
go to the Hypo indicator level set up. At the Hypo level set up, press the
left (◄) button to adjust the Hypo level. Then press the right (►) button to
go to the Test Reminder set up.
Note:
The meter allows the hypoglycemia level to be set as high as 100
mg/dL. The hypoglycemia level should be below the hyperglycemia level.
The proper hypoglycemia blood glucose level for each patient should be
determined by their physician.