AMC-DTM-Sensor Transmitter
- 23 -
6
Calibration Procedure
6.1 General
Before the unit can be calibrated, the correct sensor must be set in the sensor pre-config menu.
These are found in the menu “log in”, “options”, “advanced”, “pre-config”. Gas sensor is
calibrated using two gas concentrations: Zero concentration and an adjustable concentration
value (Span). Calibration is deemed completed when both Zero and Span points have been
calibrated. Figure 6-1 illustrates the basic equipment step-up to calibrate the AMC DTM
sensor/transmitter.
Figure 6-1 Basic equipment step-up for calibration.