AMC-DTM-Sensor Transmitter
- 6 -
For installation to an AMC Monitor, the AMC-DTM series sensor/transmitter is field wired as
illustrated in Figure 4-4. The terminal block pinout for TB-3 is given in Table 4-2 and output
specifications are provided in Table 4-3.
Figure 4-4: Typical Field Wiring of AMC-DTM to AMC Monitor
Terminal Type
Description
1 PWR
Input
Voltage
(+)
2 PWR
Input
Voltage
(+)
3
ANALOG OUT
4-20mA Output Signal (S)
4
PWR
Power Supply Common (-)
5
PWR
Power Supply Common (-)
Table 4-2: TB3 Interface Pinout
Value
Parameter
Min
Typ
Max
Units
Load Resistance
n/a
250
275
ohms
Loop Current
0.25
21
mA
Conductor Size
See
Note
16
AWG
Table 4-3: 4-20mA Output Specifications