Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 49 of 60
Special Conditions for Safe Use
•
Ambient temperature range: -40 °C
≤
T
a
≤
+60 °C
•
Verified values of the maximum gaps and minimum constructional length of flameproof joints
of this enclosure are different from relevant minimum and maximum values mentioned in
standard. To obtain information about joints dimension it is necessary to contact the
manufacturer.
•
Do not open the instrument when energised.
•
Intensive electrostatic charging processes have to be prevented.
•
Maintenance or repairment according to the type of protection "d" is only allowed by the
manufacturer.
•
IP64
–
vertical, filter unit down
Cable gland
•
The housing of the detector has to be equipped with flameproof cable gland Ex d IIC certified
according to Directive 2014/34/EU (ATEX).
•
M20 x 1,5; Torque 8
–
12 Nm
EMC Conformance according to the Directive 2014/30/EU
EN 50270:2006 Type 2, EN 61000-6-3:2007