Smart Photoionization Detector Module PIDS3
User Manual
PIDS3 - User Manual (en) - 1.03.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 43 of 58
12
Modbus Communication
The communication is compatible with the Modbus specification.
Signal level
RS485 differentials level
Parameter
115200 Baud, 57600 Baud, 34800 Baud, 19200 Baud, 9600 Baud
Modes ASCII and RTU
Device unique address
Termination resistor enabling
The default values are:
Baud rate
115200
Mode
RTU
Gas Device unique address
10
Termination resistor enable
false
USAGE
–
COMM SELECT Jumper J1
To use the MODBUS communication the jumper J1 must not be plugged.
12.1
Register map
Register 3xxxx
–
Access via function code 0x04 (read only)
12.1.1
Module information register
Name
Access
Register no
Length
Format
Device identification
RO
30001
32 byte
String
Device serial number
RO
30017
32 byte
String
Measurement config
gasname
RO
30033
16 byte
String
Measurement config
calibration method
RO
30041
16 byte
String