Calibration procedure
Action
UART command
start
pids.start
values
pids.values
write
pids.calib
verify
Included in Action ‘write’ with no
error as answer
save
pids.savedata
Invalid calibration values may be caused by:
•
MEASURE state not started
•
no clean air applied during ZERO calibration step
•
no or inappropriate calibration gas applied during SPAN calibration step
ZERO calibration data
compensated current e.g.
14.000 pA
concentration
0.0 ppm
applying SPAN gas
> 3 min
applying ZERO gas
> 3 min
SPAN calibration data
compensated current e.g.
1334.000 pA
concentration e.g.
100.0 ppm
Action ‘write’
Action ‘verify’
Action 'save’
Action ‘
values
’
Action ‘
values
’
Action ‘start’