Smart Photoionization Detector Modul PIDS3 briefly
The inlet and outlet for gas and the cable gland are located at the following positions.
The sample gas will be conducted through the sample gas inlet. A glass-fiber syringe-filter protects the
device against material pollution.
Gas Inlet
Type
male luer
Fit to tube Inner diameter
3.0 mm to 3.5 mm
Gas Outlet
Type
thread hole M5
Fit to connectors with M5 thread
E.g., SMC M5-AU-3
CAUTION
Do not remove the filter for measurement this may damage the instrument.
Gas flow
Minimum
300 ml/min
Maximum
500 ml/min
Cable gland M 16 X 1,5
Minimum outer diameter
5 mm
Maximum outer diameter
9 mm