Smart Photoionization Detector Module PIDS3
User Manual
PIDS3 - User Manual (en) - 1.03.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 18 of 58
6.3
Electrical Installation
•
Shielded cable for measuring devices is recommended.
•
All cable shields to be terminated to ground earth at one end only.
•
Water or impurities can penetrate the instrument through the cable. In hazardous areas, it is
recommended to install the cable in a loop just before entry into the instrument or to slightly bend it
to prevent water from entering.
NOTICE
The power supply is defined as 10
… 2
8 VDC.
How To
–
Install Electrical Connection
1.
Open the device.
2.
Unscrew clamping nut at the cable gland.
3.
Put clamping nut on the cable. Prepare cable for installation, enable shield to connect with cable
gland.
4.
Insert cable for connection into the device.
5.
Tighten cable gland clamping nut including shield, check that cable cannot move within the cable
gland.
6.
Connect cable to the corresponding pin numbers of the terminal.
7.
Set up enclosure lid and the screws.
WARNING
The device must be installed only in compliance with the applicable regulations, otherwise
the safe operation of the instrument is not guaranteed.