![Acelity V.A.C. Ulta Safety Information And Application Instructions Download Page 26](http://html1.mh-extra.com/html/acelity/v-a-c-ulta/v-a-c-ulta_safety-information-and-application-instructions_457655026.webp)
26
BRIDGE APPLICATION
Bridge application should be used
1)
to prevent periwound maceration with wounds that are
smaller than the central disc of the pad(s) or
2)
when there is a need to place the pad(s) away
from the wound site to prevent pressure on or around the wound.
CAUTION:
Patient’s skin condition should be carefully monitored (refer to
Precautions
, Protect
Periwound Skin section).
CAUTION:
In a
vertical
bridge placement created for moderately to highly exudating wounds,
the negative pressure at the wound site could be reduced by approximately 25 mmHg for every
foot (30.5 cm) of the bridge. Consider adjusting the negative pressure setting accordingly.
USING V.A.C. VERAFLO CLEANSE™ DRESSING
2.5
cm
3
cm
Fig. 27
Fig. 33
Fig. 30
Fig. 28
Fig. 31
Fig. 34
Fig. 32
Fig. 29
Fig. 35
Fig. 36
Cut Bridge to shape
Central
disc
Bridge area
Summary of Contents for V.A.C. Ulta
Page 2: ......
Page 29: ...29 ...
Page 32: ......
Page 59: ...59 ...
Page 62: ......
Page 89: ...89 ...
Page 92: ......
Page 119: ...119 ...
Page 122: ......
Page 149: ...149 ...
Page 152: ......
Page 179: ...179 ...
Page 182: ......
Page 209: ...209 ...
Page 212: ......
Page 239: ...239 ...
Page 242: ......
Page 269: ...269 ...
Page 272: ......
Page 299: ...299 ...
Page 302: ......
Page 329: ...329 ...
Page 332: ......
Page 359: ...359 ...
Page 362: ......
Page 389: ...389 ...
Page 395: ...395 ...