ACD CAL-101 Instruction Manual Download Page 6

II. START-UP

To start the generator, insert the generating cell. In the bump station 

configuration, press the button protruding from the bottom of the cell.

The RED LEDs will light to indicate battery life remaining. All 10 LEDs 

indicate fully charged batteries. If only one RED LED is observed, the 

batteries are near depletion. During the battery test time, current is supplied 

to the generating cell and the air pump is not turned on in an attempt to speed 

up equilibrium time.  This helps to saturate the generating cell with gas. 

After the battery life indication, the pump will start to pump air through the 

instrument and the GREEN LEDs will begin to flash approximately once 

every 4 seconds. The flashing GREEN LEDs will indicate the time remaining 

in the generating cell as follows: 

 

10 LEDs lit indicates that 2 hours remain on the cell.

    1 LED lit indicates less than 12 minutes of calibration time remain.

These are the normal operating conditions for the CAL-101. Complete 

equilibrium may take an additional two to five minutes, depending upon the 

type and amount of gas being generated and the operating conditions of 

the instrument.  In the bump station configuration the unit will automatically 

shut off after thirty seconds. 

When the generating cell is inserted, if all 10 green LEDs come on constant, 

this indicates a cell failure. Remove the cell and wait for all LED’s to turn off, 

then re-insert ensuring that the cell is fully inserted correctly and properly 

aligned.

If the GREEN LEDs ‘ramp’ up and down from max to min this indicates that 

the cell life is depleted and a new generating cell should be purchased. 

At any time the instrument may be turned off by removing the generating 

cell.  The GREEN LEDs will all light indicating cell failure and the instrument 

will then shut off.

Page 4

Summary of Contents for CAL-101

Page 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Page 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Page 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Page 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Page 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Page 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Page 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Page 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Page 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Page 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Page 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Page 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Page 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Page 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Page 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Reviews: