ACD CAL-101 Instruction Manual Download Page 9

If the cell has been depleted, it should be removed and replaced with 

a new generating cell.  Please refer to the section on cell life for proper 

instructions on replacing the generating cells.  If the cell has not been 

depleted, there may be a faulty electrical connection between the cell and 

the instrument.  Remove the cell and inspect the pins protruding from the 

pump board and the corresponding pads on the cell board. Remove any 

debris and or corrosion that may be present and reinsert the cell ensuring 

proper alignment.  If either the pins  become damaged or non-repairable, 

the pump board should be replaced..

The  rate  of  airflow  through  the  instrument  is  critical  for  accurate  gas 

concentrations. The flow should be adjusted to 0.5 LPM (liters per minute) 

STP (standard temperature and pressure) of sea level and 20

o

C. (see 

chart in section VI) To trim the flow hold the magnetic screw driver over 

the 'flow trim' area of the instrument indicated with a magnet symbol. The 

GREEN LEDs will begin flashing at a once per second rate indicating the 

relative power to the pump. As you hold the magnet over this area the pump 

will continue to increase in speed (The LEDs will ramp up indicating the 

relative power) until the magnet is removed. When the magnet is reapplied 

to the area, the pump speed will decrease (the LEDs will also ramp down 

indicating the relative power) until the magnet is again removed. Each time 

the magnet is removed and reapplied the flow will change from increasing to 

decreasing. This is how the flow rate of the instrument should be adjusted 

to obtain 0.5LPM STP. Once the flow is adjusted the instrument will save 

the flow data and return to that flow rate for future use

Page 7

Summary of Contents for CAL-101

Page 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Page 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Page 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Page 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Page 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Page 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Page 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Page 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Page 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Page 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Page 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Page 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Page 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Page 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Page 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Reviews: