ACD CAL-101 Instruction Manual Download Page 11

VI. CORRECTIONS TO VOLUMETRIC FLOW

Changes in altitude and temperature can cause changes in air density, 

thus affecting the output of the CAL-101. The output of the CAL-101 was 

calibrated to STP (standard temperature and pressure) of sea level and 

20°C.  To compensate for differences in altitude or temperature from STP, 

the following tables should be utilized:

Note:  Altitude and Temperature are cumulative, i.e., you must correct 

for both factors to obtain the most accurate calibration.

  

Altitude Correction Table

Temperature Correction Table

Altitude

Corrected Out-

put (multiply by)

Temperature 

(°C)

Corrected (multiply by)

0

1.000

-30

0.829

1000

1.038

-20

0.863

2000

1.076

-10

0.898

3000

1.116

0

0.932

4000

1.158

10

0.966

5000

1.202

20

1.000

6000

1.248

30

1.034

40

1.068

50

1.102

Example:

At 3200 feet with a temperature of 29°C.  The corrected  volumetric flow 

rate is

 

.5 LPM  x 1.116 x 1.034 =  .577 LPM (volumetric) 

Page 9

Summary of Contents for CAL-101

Page 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Page 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Page 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Page 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Page 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Page 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Page 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Page 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Page 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Page 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Page 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Page 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Page 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Page 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Page 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Reviews: